Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 5700A: (90 day) CAL VER ACV IEEE /5790 DATE: 26-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 2.9 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 61 NUMBER OF LINES: 376 CONFIGURATION: Fluke 5790A ============================================================================= # # Source: # Fluke 5700A Service Manual, Section 3, Rev 2, May 1991 # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # Sub Wait for Operation Complete (OPC) IEEE # # Required Files: # 5790CON1.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5790A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5790A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R Q N F W # Set "Wait for OPC" timeout to 30000ms. 1.002 MATH M[88] = 30000 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this verification. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 OPBR Is the Fluke 5700A to be calibrated a configured system 1.005 OPBR instrument connected to IEEE-488 port 0 on this PC? 1.006 JMPT 1.011 1.007 DISP Connect an IEEE-488 cable from the 5700A to 488 Port 1. 1.008 DISP Turn on the 5700A and allow 30 minutes warmup time. 1.009 DISP Allow the Fluke 5700A to stablize in an environment 1.009 DISP with ambient temperature of 21 to 25 C and relative 1.009 DISP humidity of less than 70%. 1.010 JMP 1.012 1.011 DISP Enter the address of the 5700A (MET/CAL default = 15) 1.011 DISP when prompted for the UUT address. 1.012 IEEE *CLS;*RST 1.013 PIC 5790CON1 1.014 IEEE EXTGUARD ON 1.015 RSLT =ACV CALIBRATION 1.016 RSLT = 1.017 ACMS 2mV M 1.018 HEAD ACV CALIBRATION {2.2mV RANGE}: 2mV @ 10Hz 1.019 IEEE [T3000]OUT 2mV,10Hz;OPER 1.020 CALL Sub Wait for Operation Complete (OPC) IEEE 1.021 JMPL END M[89] <= 0 1.022 5790 2.2 2.000000mV 0.305% 10H I1 2W 2.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 20Hz 2.002 IEEE [T3000]OUT 2mV,20Hz;OPER 2.003 CALL Sub Wait for Operation Complete (OPC) IEEE 2.004 JMPL END M[89] <= 0 2.005 5790 2.2 2.000000mV 0.272% 20H I1 2W 3.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 40Hz 3.002 IEEE [T3000]OUT 2mV,40Hz;OPER 3.003 CALL Sub Wait for Operation Complete (OPC) IEEE 3.004 JMPL END M[89] <= 0 3.005 5790 2.2 2.000000mV 0.26% 40H I1 2W 4.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 1kHz 4.002 IEEE [T3000]OUT 2mV,1kHz;OPER 4.003 CALL Sub Wait for Operation Complete (OPC) IEEE 4.004 JMPL END M[89] <= 0 4.005 5790 2.2 2.000000mV 0.26% 1kH I1 2W 5.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 20kHz 5.002 IEEE [T3000]OUT 2mV,20kHz;OPER 5.003 CALL Sub Wait for Operation Complete (OPC) IEEE 5.004 JMPL END M[89] <= 0 5.005 5790 2.2 2.000000mV 0.26% 20kH I1 2W 6.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 50kHz 6.002 IEEE [T3000]OUT 2mV,50kHz;OPER 6.003 CALL Sub Wait for Operation Complete (OPC) IEEE 6.004 JMPL END M[89] <= 0 6.005 5790 2.2 2.000000mV 0.29% 50kH I1 2W 7.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 100kHz 7.002 IEEE [T3000]OUT 2mV,100kHz;OPER 7.003 CALL Sub Wait for Operation Complete (OPC) IEEE 7.004 JMPL END M[89] <= 0 7.005 5790 2.2 2.000000mV 0.49% 100kH I1 2W 8.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 300kHz 8.002 IEEE [T3000]OUT 2mV,300kHz;OPER 8.003 CALL Sub Wait for Operation Complete (OPC) IEEE 8.004 JMPL END M[89] <= 0 8.005 5790 2.2 2.000000mV 0.87% 300kH I1 2W 9.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 500kHz 9.002 IEEE [T3000]OUT 2mV,500kHz;OPER 9.003 CALL Sub Wait for Operation Complete (OPC) IEEE 9.004 JMPL END M[89] <= 0 9.005 5790 2.2 2.000000mV 1.67% 500kH I1 2W 10.001 HEAD ACV CALIBRATION 2.2mV RANGE: 2mV @ 1MHz 10.002 IEEE [T3000]OUT 2mV,1MHz;OPER 10.003 CALL Sub Wait for Operation Complete (OPC) IEEE 10.004 JMPL END M[89] <= 0 10.005 5790 2.2 2.000000mV 1.83% 1MH I1 2W 11.001 RSLT = 11.002 ACMS 20mV M 11.003 HEAD ACV CALIBRATION {22mV RANGE}: 20mV @ 10Hz 11.004 IEEE [T3000]OUT 20mV,10Hz;OPER 11.005 CALL Sub Wait for Operation Complete (OPC) IEEE 11.006 JMPL END M[89] <= 0 11.007 5790 22 20.00000mV 850P% 10H I1 2W 12.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 20Hz 12.002 IEEE [T3000]OUT 20mV,20Hz;OPER 12.003 CALL Sub Wait for Operation Complete (OPC) IEEE 12.004 JMPL END M[89] <= 0 12.005 5790 22 20.00000mV 520P% 20H I1 2W 13.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 40Hz 13.002 IEEE [T3000]OUT 20mV,40Hz;OPER 13.003 CALL Sub Wait for Operation Complete (OPC) IEEE 13.004 JMPL END M[89] <= 0 13.005 5790 22 20.00000mV 410P% 40H I1 2W 14.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 1kHz 14.002 IEEE [T3000]OUT 20mV,1kHz;OPER 14.003 CALL Sub Wait for Operation Complete (OPC) IEEE 14.004 JMPL END M[89] <= 0 14.005 5790 22 20.00000mV 410P% 1kH I1 2W 15.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 20kHz 15.002 IEEE [T3000]OUT 20mV,20kHz;OPER 15.003 CALL Sub Wait for Operation Complete (OPC) IEEE 15.004 JMPL END M[89] <= 0 15.005 5790 22 20.00000mV 410P% 20kH I1 2W 16.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 50kHz 16.002 IEEE [T3000]OUT 20mV,50kHz;OPER 16.003 CALL Sub Wait for Operation Complete (OPC) IEEE 16.004 JMPL END M[89] <= 0 16.005 5790 22 20.00000mV 670P% 50kH I1 2W 17.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 100kHz 17.002 IEEE [T3000]OUT 20mV,100kHz;OPER 17.003 CALL Sub Wait for Operation Complete (OPC) IEEE 17.004 JMPL END M[89] <= 0 17.005 5790 22 20.00000mV 0.13% 100kH I1 2W 18.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 300kHz 18.002 IEEE [T3000]OUT 20mV,300kHz;OPER 18.003 CALL Sub Wait for Operation Complete (OPC) IEEE 18.004 JMPL END M[89] <= 0 18.005 5790 22 20.00000mV 0.20% 300kH I1 2W 19.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 500kHz 19.002 IEEE [T3000]OUT 20mV,500kHz;OPER 19.003 CALL Sub Wait for Operation Complete (OPC) IEEE 19.004 JMPL END M[89] <= 0 19.005 5790 22 20.00000mV 0.32% 500kH I1 2W 20.001 HEAD ACV CALIBRATION 22mV RANGE: 20mV @ 1MHz 20.002 IEEE [T3000]OUT 20mV,1MHz;OPER 20.003 CALL Sub Wait for Operation Complete (OPC) IEEE 20.004 JMPL END M[89] <= 0 20.005 5790 22 20.00000mV 0.48% 1MH I1 2W 21.001 RSLT = 21.002 ACMS 200mV M 21.003 HEAD ACV CALIBRATION {220mV RANGE}: 200mV @ 10Hz 21.004 IEEE [T3000]OUT 200mV,10Hz;OPER 21.005 CALL Sub Wait for Operation Complete (OPC) IEEE 21.006 JMPL END M[89] <= 0 21.007 5790 220 200.0000mV 630P% 10H I1 2W 22.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 20Hz 22.002 IEEE [T3000]OUT 200mV,20Hz;OPER 22.003 CALL Sub Wait for Operation Complete (OPC) IEEE 22.004 JMPL END M[89] <= 0 22.005 5790 220 200.0000mV 270P% 20H I1 2W 23.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 40Hz 23.002 IEEE [T3000]OUT 200mV,40Hz;OPER 23.003 CALL Sub Wait for Operation Complete (OPC) IEEE 23.004 JMPL END M[89] <= 0 23.005 5790 220 200.0000mV 150P% 40H I1 2W 24.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 1kHz 24.002 IEEE [T3000]OUT 200mV,1kHz;OPER 24.003 CALL Sub Wait for Operation Complete (OPC) IEEE 24.004 JMPL END M[89] <= 0 24.005 5790 220 200.0000mV 150P% 1kH I1 2W 25.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 20kHz 25.002 IEEE [T3000]OUT 200mV,20kHz;OPER 25.003 CALL Sub Wait for Operation Complete (OPC) IEEE 25.004 JMPL END M[89] <= 0 25.005 5790 220 200.0000mV 150P% 20kH I1 2W 26.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 50kHz 26.002 IEEE [T3000]OUT 200mV,50kHz;OPER 26.003 CALL Sub Wait for Operation Complete (OPC) IEEE 26.004 JMPL END M[89] <= 0 26.005 5790 220 200.0000mV 380P% 50kH I1 2W 27.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 100kHz 27.002 IEEE [T3000]OUT 200mV,100kHz;OPER 27.003 CALL Sub Wait for Operation Complete (OPC) IEEE 27.004 JMPL END M[89] <= 0 27.005 5790 220 200.0000mV 950P% 100kH I1 2W 28.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 300kHz 28.002 IEEE [T3000]OUT 200mV,300kHz;OPER 28.003 CALL Sub Wait for Operation Complete (OPC) IEEE 28.004 JMPL END M[89] <= 0 28.005 5790 220 200.0000mV 0.12% 300kH I1 2W 29.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 500kHz 29.002 IEEE [T3000]OUT 200mV,500kHz;OPER 29.003 CALL Sub Wait for Operation Complete (OPC) IEEE 29.004 JMPL END M[89] <= 0 29.005 5790 220 200.0000mV 0.19% 500kH I1 2W 30.001 HEAD ACV CALIBRATION 220mV RANGE: 200mV @ 1MHz 30.002 IEEE [T3000]OUT 200mV,1MHz;OPER 30.003 CALL Sub Wait for Operation Complete (OPC) IEEE 30.004 JMPL END M[89] <= 0 30.005 5790 220 200.0000mV 0.38% 1MH I1 2W 31.001 RSLT = 31.002 ACMS 2V M 31.003 HEAD ACV CALIBRATION {2.2V RANGE}: 2V @ 10Hz 31.004 IEEE [T3000]OUT 2V,10Hz;OPER 31.005 CALL Sub Wait for Operation Complete (OPC) IEEE 31.006 JMPL END M[89] <= 0 31.007 5790 2.2 2.000000V 600P% 10H I1 2W 32.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 20Hz 32.002 IEEE [T3000]OUT 2V,20Hz;OPER 32.003 CALL Sub Wait for Operation Complete (OPC) IEEE 32.004 JMPL END M[89] <= 0 32.005 5790 2.2 2.000000V 185P% 20H I1 2W 33.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 40Hz 33.002 IEEE [T3000]OUT 2V,40Hz;OPER 33.003 CALL Sub Wait for Operation Complete (OPC) IEEE 33.004 JMPL END M[89] <= 0 33.005 5790 2.2 2.000000V 78.5P% 40H I1 2W 34.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 1kHz 34.002 IEEE [T3000]OUT 2V,1kHz;OPER 34.003 CALL Sub Wait for Operation Complete (OPC) IEEE 34.004 JMPL END M[89] <= 0 34.005 5790 2.2 2.000000V 78.5P% 1kH I1 2W 35.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 20kHz 35.002 IEEE [T3000]OUT 2V,20kHz;OPER 35.003 CALL Sub Wait for Operation Complete (OPC) IEEE 35.004 JMPL END M[89] <= 0 35.005 5790 2.2 2.000000V 78.5P% 20kH I1 2W 36.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 50kHz 36.002 IEEE [T3000]OUT 2V,50kHz;OPER 36.003 CALL Sub Wait for Operation Complete (OPC) IEEE 36.004 JMPL END M[89] <= 0 36.005 5790 2.2 2.000000V 140P% 50kH I1 2W 37.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 100kHz 37.002 IEEE [T3000]OUT 2V,100kHz;OPER 37.003 CALL Sub Wait for Operation Complete (OPC) IEEE 37.004 JMPL END M[89] <= 0 37.005 5790 2.2 2.000000V 290P% 100kH I1 2W 38.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 300kHz 38.002 IEEE [T3000]OUT 2V,300kHz;OPER 38.003 CALL Sub Wait for Operation Complete (OPC) IEEE 38.004 JMPL END M[89] <= 0 38.005 5790 2.2 2.000000V 515P% 300kH I1 2W 39.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 500kHz 39.002 IEEE [T3000]OUT 2V,500kHz;OPER 39.003 CALL Sub Wait for Operation Complete (OPC) IEEE 39.004 JMPL END M[89] <= 0 39.005 5790 2.2 2.000000V 0.13% 500kH I1 2W 40.001 HEAD ACV CALIBRATION 2.2V RANGE: 2V @ 1MHz 40.002 IEEE [T3000]OUT 2V,1MHz;OPER 40.003 CALL Sub Wait for Operation Complete (OPC) IEEE 40.004 JMPL END M[89] <= 0 40.005 5790 2.2 2.000000V 0.27% 1MH I1 2W 41.001 RSLT = 41.002 ACMS 20V M 41.003 HEAD ACV CALIBRATION {22V RANGE}: 20V @ 10Hz 41.004 IEEE [T3000]OUT 20V,10Hz;OPER 41.005 CALL Sub Wait for Operation Complete (OPC) IEEE 41.006 JMPL END M[89] <= 0 41.007 5790 22 20.00000V 600P% 10H I1 2W 42.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 20Hz 42.002 IEEE [T3000]OUT 20V,20Hz;OPER 42.003 CALL Sub Wait for Operation Complete (OPC) IEEE 42.004 JMPL END M[89] <= 0 42.005 5790 22 20.00000V 185P% 20H I1 2W 43.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 40Hz 43.002 IEEE [T3000]OUT 20V,40Hz;OPER 43.003 CALL Sub Wait for Operation Complete (OPC) IEEE 43.004 JMPL END M[89] <= 0 43.005 5790 22 20.00000V 78.5P% 40H I1 2W 44.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 1kHz 44.002 IEEE [T3000]OUT 20V,1kHz;OPER 44.003 CALL Sub Wait for Operation Complete (OPC) IEEE 44.004 JMPL END M[89] <= 0 44.005 5790 22 20.00000V 78.5P% 1kH I1 2W 45.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 20kHz 45.002 IEEE [T3000]OUT 20V,20kHz;OPER 45.003 CALL Sub Wait for Operation Complete (OPC) IEEE 45.004 JMPL END M[89] <= 0 45.005 5790 22 20.00000V 78.5P% 20kH I1 2W 46.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 50kHz 46.002 IEEE [T3000]OUT 20V,50kHz;OPER 46.003 CALL Sub Wait for Operation Complete (OPC) IEEE 46.004 JMPL END M[89] <= 0 46.005 5790 22 20.00000V 140P% 50kH I1 2W 47.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 100kHz 47.002 IEEE [T3000]OUT 20V,100kHz;OPER 47.003 CALL Sub Wait for Operation Complete (OPC) IEEE 47.004 JMPL END M[89] <= 0 47.005 5790 22 20.00000V 270P% 100kH I1 2W 48.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 300kHz 48.002 IEEE [T3000]OUT 20V,300kHz;OPER 48.003 CALL Sub Wait for Operation Complete (OPC) IEEE 48.004 JMPL END M[89] <= 0 48.005 5790 22 20.00000V 635P% 300kH I1 2W 49.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 500kHz 49.002 IEEE [T3000]OUT 20V,500kHz;OPER 49.003 CALL Sub Wait for Operation Complete (OPC) IEEE 49.004 JMPL END M[89] <= 0 49.005 5790 22 20.00000V 0.155% 500kH I1 2W 50.001 HEAD ACV CALIBRATION 22V RANGE: 20V @ 1MHz 50.002 IEEE [T3000]OUT 20V,1MHz;OPER 50.003 CALL Sub Wait for Operation Complete (OPC) IEEE 50.004 JMPL END M[89] <= 0 50.005 5790 22 20.00000V 0.325% 1MH I1 2W 51.001 RSLT = 51.002 ACMS 200V M 51.003 HEAD ACV CALIBRATION {220V RANGE}: 200V @ 10Hz 51.004 IEEE [T3000]OUT 200V,10Hz;*CLS;OPER 51.005 CALL Sub Wait for Operation Complete (OPC) IEEE 51.006 JMPL END M[89] <= 0 51.007 5790 220 200.0000V 600P% 10H I1 2W 52.001 HEAD ACV CALIBRATION 220V RANGE: 200V @ 20Hz 52.002 IEEE [T3000]OUT 200V,20Hz;OPER 52.003 CALL Sub Wait for Operation Complete (OPC) IEEE 52.004 JMPL END M[89] <= 0 52.005 5790 220 200.0000V 185P% 20H I1 2W 53.001 HEAD ACV CALIBRATION 220V RANGE: 200V @ 40Hz 53.002 IEEE [T3000]OUT 200V,40Hz;OPER 53.003 CALL Sub Wait for Operation Complete (OPC) IEEE 53.004 JMPL END M[89] <= 0 53.005 5790 220 200.0000V 85P% 40H I1 2W 54.001 HEAD ACV CALIBRATION 220V RANGE: 200V @ 1kHz 54.002 IEEE [T3000]OUT 200V,1kHz;OPER 54.003 CALL Sub Wait for Operation Complete (OPC) IEEE 54.004 JMPL END M[89] <= 0 54.005 5790 220 200.0000V 85P% 1kH I1 2W 55.001 HEAD ACV CALIBRATION 220V RANGE: 200V @ 20kHz 55.002 IEEE [T3000]OUT 200V,20kHz;OPER 55.003 CALL Sub Wait for Operation Complete (OPC) IEEE 55.004 JMPL END M[89] <= 0 55.005 5790 220 200.0000V 85P% 20kH I1 2W 56.001 HEAD ACV CALIBRATION 220V RANGE: 200V @ 50kHz 56.002 IEEE [T3000]OUT 200V,50kHz;OPER 56.003 CALL Sub Wait for Operation Complete (OPC) IEEE 56.004 JMPL END M[89] <= 0 56.005 5790 220 200.0000V 240P% 50kH I1 2W 57.001 HEAD ACV CALIBRATION 220V RANGE: 200V @ 100kHz 57.002 IEEE [T3000]OUT 200V,100kHz;OPER 57.003 CALL Sub Wait for Operation Complete (OPC) IEEE 57.004 JMPL END M[89] <= 0 57.005 5790 220 200.0000V 600P% 100kH I1 2W 58.001 RSLT = 58.002 ACMS 1000V M 58.003 HEAD ACV CALIBRATION {1100V RANGE}: 1000V @ 50Hz 58.004 IEEE [T3000]OUT 1000V,50Hz;BOOST OFF;OPER 58.005 CALL Sub Wait for Operation Complete (OPC) IEEE 58.006 JMPL END M[89] <= 0 58.007 5790 1000 1000.000V 84P% 50H I1 2W 59.001 HEAD ACV CALIBRATION 1100V RANGE: 1000V @ 300Hz 59.002 IEEE [T3000]OUT 1000V,300Hz;BOOST OFF;OPER 59.003 CALL Sub Wait for Operation Complete (OPC) IEEE 59.004 JMPL END M[89] <= 0 59.005 5790 1000 1000.000V 84P% 300H I1 2W 60.001 HEAD ACV CALIBRATION 1100V RANGE: 1000V @ 1kHz 60.002 IEEE [T3000]OUT 1000V,1kHz;BOOST OFF;OPER 60.003 CALL Sub Wait for Operation Complete (OPC) IEEE 60.004 JMPL END M[89] <= 0 60.005 5790 1000 1000.000V 84P% 1kH I1 2W 61.001 LABEL END 61.002 IEEE [T0]*RST 61.003 5790 * S 61.004 END